Thirteen-year Experience of Permanent Epicardial Pacing in Children.
- Author:
Jeong Ryul LEE
1
;
Kook Nam HAN
;
Hong Gook LIM
;
Woong Han KIM
;
Yong Jin KIM
;
Joon Ryang RHO
;
Eun Jung BAE
;
Chung Il NOH
;
Yong Soo YOON
;
Curie AHN
Author Information
1. Department of Thoracic and Cardiovascular Surgery, Seoul National University Children's Hospital, Seoul National University College of Medicine, Korea. jrl@plaza.snu.ac.kr
- Publication Type:Original Article
- Keywords:
Pacemaker, artificial;
Congenital heart defect
- MeSH:
Child*;
Heart Defects, Congenital;
Humans;
Longevity;
Pacemaker, Artificial;
Reoperation
- From:The Korean Journal of Thoracic and Cardiovascular Surgery
2004;37(6):499-503
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
BACKGROUND: We investigated the longevity, thresholds of epicardial pacemaker and causes of reoperation in the pediatric patients who underwent epicardial pacemaker implantation performed during the last 13 years MATERIAL AND METHOD: 121 operations were performed in 83 patients from January 1989 to July 2002. We analyzed the stimulation threshold, resistance, R-wave and P-wave, and sensitivity of pacemaker lead at initial implantation. Longevity and causes of reoperations were investigated. RESULT: At implantation, epicardial ventricular mean stimulation threshold was 1.2+/-0.1 (0.1~5) mV, mean resistance was 519.1+/-18.1 (319~778) Ohm, and mean R-wave sensitivity was 8.9+/-0.7 (4~20) mV, and mean P wave sensivity was 2.5+/-0.7 (0.4~12) mV. The mean longevity of pacemaker generator was 64.7+/-3.7 (2~196) months. The reoperation free rate was 94.6% for 1 year, 93.6% for 2 years, 80.8% for 5years, 63.7% for 7 years, and 45.5% for 10 years. The causes of reoperation were battery waste in 26 cases and lead malfunction in 9 cases. There was no postoperative death related to pacemaker malfunction. CONCLUSION: In the childrens, average longevity of epicardial pacemaker was within acceptable range. 19.1% of the patients required pacemaker related reoperation. However, recent developments, including steroid eluting lead, 6.7% of the patients required pacemaker related reoperation, look promising in expansion of pacemaker life span.