The Production and Evaluation of the Tissue-equivalent Phantom for the Magnetic Resonance Imaging.
10.3348/jkrs.1994.30.6.1151
- Author:
Young Hoon RYU
;
Jae Hyun CHO
;
Jin Suck SUH
;
Jae Myun LEE
;
Eun Kee JEONG
- Publication Type:Original Article
- MeSH:
Agar;
Humans;
Magnetic Resonance Imaging*;
Polysorbates;
Relaxation
- From:Journal of the Korean Radiological Society
1994;30(6):1151-1155
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: For the production and evaluation of the tissue-equivalent phantom. MATERIALS AND METHODS: We used agarous gel and oil as a basic component of the mixture and added Tween 80 for the stabilization of phantoms. We did the test for homogeneity and measured T1 and T2 relexation times of each phantom tube. RESULTS: T1 relaxation time ranged from 642 to 2781 msec and T2 relaxation times from 42 to 157 msec. Each phantom was significantly different in T1 relaxation time and T2 relaxation time (p < .0001). CONCLUSION: Tissue equivalent phantom may provide good information on the optimal sequence before MR imaging of patients and may be valuable if it is used with the patients' MR imaging.