Measurement of White-to-White Diameter and Anterior Chamber Depth by Dual Scheimpflug Camera.
10.3341/jkos.2010.51.2.169
- Author:
Kyung Eun HAN
1
;
Roo Min JUN
Author Information
1. Department of Ophthalmology, Ewha Womans University School of Medicine, Seoul, Korea. jrmoph@ewha.ac.kr
- Publication Type:Original Article
- Keywords:
Anterior chamber depth;
GALILEI;
Scheimpflug camera;
Sulcus diameter;
White-to-white diameter
- MeSH:
Anterior Chamber;
Eye;
Microscopy, Acoustic;
Utah
- From:Journal of the Korean Ophthalmological Society
2010;51(2):169-174
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: To estimate the horizontal and vertical white-to-white diameters (WTW) and anterior chamber depths (ACD) with a dual Scheimpflug camera (GALILEI(TM), Ziemer, Switzerland) and to compare the estimates measured by a measuring caliper and ultrasound biomicroscopy (UBM PlusTM, Paradigm Inc., Utah, USA) in normal subjects. METHODS: Forty-four eyes of 23 subjects were evaluated. Corneal diameter as measured by GALILEI was directly compared with the white-to-white diameter (WTW) measured by a caliper and the correlation with ciliary sulcus diameter (STS) by UBM was evaluated. The anterior chamber depth (ACD) as measured by GALILEI(TM) was compared with the estimates measured by UBM. RESULTS: The horizontal and vertical diameters of WTW by GALILEI were not significantly different from the measurements taken by calipers (p>0.05, p>0.05, respectively), and there were no correlations between the measurements (r=-0.074, p>0.05 at 180degrees r=0.297, p>0.05 at 90degrees, respectively). The estimates by GALILEI did not correlate with those measured with UBM (r=-0.006, p>0.05 at 180degrees r=-0.222, p>0.05 at 90degrees, respectively). However, the mean ACD by GALILEI was deeper than thatby UBM (p<0.01), and the measurements correlated strongly with each other (r=0.760; p<0.01). CONCLUSIONS: The mean WTW measured by GALILEI was not significantly different from the measurements taken by calipers and the measurements did not correlate with each other. There was also no correlation with the measurementsby GALILEI and UBM. ACD by GALILEI was measured to be deeper those that by UBM.