Efficacy of a Respiratory Training System on the Regularity of Breathing.
10.3857/jkstro.2008.26.3.181
- Author:
Eunhyuk SHIN
1
;
Hee Chul PARK
;
Youngyih HAN
;
Sang Gyu JU
;
Jung Suk SHIN
;
Yong Chan AHN
Author Information
1. Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. rophc@skku.edu
- Publication Type:Original Article
- Keywords:
4-dimensional radiation therapy;
Respiratory gating radiation therapy;
Breathing cycle
- From:The Journal of the Korean Society for Therapeutic Radiology and Oncology
2008;26(3):181-188
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: In order to enhance the efficiency of respiratory gated 4-dimensional radiation therapy for more regular and stable respiratory period and amplitude, a respiration training system was designed, and its efficacy was evaluated. MATERIALS AND METHODS: The experiment was designed to measure the difference in respiration regularity following the use of a training system. A total of 11 subjects (9 volunteers and 2 patients) were included in the experiments. Three different breathing signals, including free breathing (free-breathing), guided breathing that followed training software (guided-breathing), and free breathing after the guided-breathing (post guided-breathing), were consecutively recorded in each subject. The peak-to-peak (PTP) period of the breathing signal, standard deviation (SD), peak-amplitude and its SD, area of the one cycle of the breathing wave form, and its root mean square (RMS) were measured and computed. RESULTS: The temporal regularity was significantly improved in guided-breathing since the SD of breathing period reduced (free-breathing 0.568 vs guided-breathing 0.344, p=0.0013). The SD of the breathing period representing the post guided-breathing was also reduced, but the difference was not statistically significant (free-breathing 0.568 vs. guided-breathing 0.512, p=ns). Also the SD of measured amplitude was reduced in guided-breathing (free-breathing 1.317 vs. guided-breathing 1.068, p=0.187), although not significant. This indicated that the tidal volume for each breath was kept more even in guided-breathing compared to free-breathing. There was no change in breathing pattern between free-breathing and guided-breathing. The average area of breathing wave form and its RMS in postguided-breathing, however, was reduced by 7% and 5.9%, respectively. CONCLUSION: The guided-breathing was more stable and regular than the other forms of breathing data. Therefore, the developed respiratory training system was effective in improving the temporal regularity and maintaining a more even tidal volume.