Left Ventricular Ejection Fraetion Determmed by Cated Tl-201 Perfusion SPECT and Quantitative Software.
- Author:
In Young HYUN
;
Sung Eun KIM
;
Jeong Kee SEO
;
Eui Soo HONG
;
Jun KWAN
;
Keum Soo PARK
;
Woo Hyung LEE
- Publication Type:Original Article
- Keywords:
Left ventricular ejection fraction;
TI-201;
Gated myocardial perfusion SPECT
- MeSH:
Perfusion*;
Tomography, Emission-Computed, Single-Photon*
- From:Korean Journal of Nuclear Medicine
2000;34(3):222-227
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: We compared estimates of ejection fraction (EF) determined by gated Tl-201 perfusion SPECT (g-TI-SPECT) with those by gated blood pool (GBP) scan. MATERIALS AND METHODS: Eighteen subjects underwent g-TI-SPECT and GBP scan. After reconstruction of g-TI-SPECT, we measured EF with Cedars software. The comparison of the EF with g-TI-SPECT and GHP scan was assessed by correlation analysis and Bland Altman plot. RESULTS: The estimates of EF were significantly different (p<0.05) with g-TI-SPECT (40%+/-14%) and GBP scan (43%+/-14%). There was an excellent correlation of EF between e-TI-SPECT and GBP scan (r=0.94, p<0.001). The mean difference of EF between GRP scan and g-TI- SPECT was +3.2%, Ninety-five percent limits of agreement were +9,8%. EF between g-TI-SPECT and GBP scan were in poor agreement. CONCLUSION: The estimates of EF by g-TI-SPECT was well correlated with those by GBP scan. However, EF of g-TI-SPECT doesn't agree with EF of GBP scan. EF of g-TI-SPECT cant be used interchangeably with EF of GBP scan.