Performance Evaluation of BAROZEN H, a Networking Blood Glucose Monitoring System for Medical Institutions.
- Author:
Jun Hyung LEE
1
;
Young Joo CHA
Author Information
- Publication Type:Original Article
- Keywords: Blood glucose; Blood glucose self-monitoring; Diabetes mellitus; Glucose oxidase; Hematocrit; Point-of-care systems
- MeSH: Blood Glucose Self-Monitoring; Blood Glucose*; Chemistry; Diabetes Mellitus; Glucose Oxidase; Hematocrit; Point-of-Care Systems
- From:Laboratory Medicine Online 2015;5(2):69-76
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: We evaluated the analytical performance of Barozen H (i-SENS Inc., Korea), a new glucometer equipped with networking function for medical institutions, according to the ISO 15197:2003 and ISO/DIS 15197:2011 guidelines. METHODS: We measured the precision of 10 Barozen H glucometers, in terms of repeatability and intermediate precision, and determined their accuracy relative to that of automatic chemistry analyzer AU5421 (Beckman Coulter, USA). Three other glucometers-Precision PCx (Abbott, USA), Glucocard Sigma (Arkray, Japan), and SureStep Flexx (Johnson & Johnson, USA) were also evaluated, and their accuracies and hematocrit interferences were compared. RESULTS: The standard deviation and coefficient of variation of Barozen H for repeatability and intermediate precision were 0.11-0.15 mmol/L and 2.3-3.6%, respectively. With respect to accuracy, in accordance with ISO 15197:2003 criteria, Barozen H yielded 98.0% of results within +/-0.83 mmol/L or +/-20%. Further, per the ISO/DIS 15197:2011 criteria, 95.2% of results were within +/-0.83 mmol/L or +/-15%; Barozen H was the only glucometer satisfying the more stringent ISO/DIS 15197:2011 criteria. Error grid analysis showed that all results from Barozen H were in zone A, indicating its excellent clinical accuracy. Hematocrit, ranging from 20% to 60% did not cause any significant interference. CONCLUSIONS: Barozen H showed excellent analytical performance, and it was the most clinically accurate glucometer tested. It can be expected to provide reliable results satisfying ISO/DIS 15197:2011 as well as ISO 15197:2003 criteria.