Three Cases of Hemolysis After Transcatheter Closure of A Patent Ductus Arteriosus.
- Author:
Sang Bum LEE
1
;
Myung Chul HYUN
Author Information
1. Department of Pediatrics, Kyungpook National University, College of Medicine, Taegu, Korea.
- Publication Type:Case Report
- Keywords:
Patent ductus arteriosus;
Transcatheter occlusion;
Hemolysis
- MeSH:
Ductus Arteriosus, Patent*;
Hemolysis*;
Humans;
Ligation
- From:Journal of the Korean Pediatric Society
2000;43(10):1399-1403
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
Severe hemolysis occurred after transcatheter occlusion of a patent ductus arteriosus in 3 patients among 41 patients who underwent transcatheter occlusion of a patent ductus arteriosus. The problems were managed by a standard surgical ligation of the duct, leaving the occlusion device(the Rashkind umbrella and a detachable coil) in situ in one patient, surgical removal of the detachable coil and division of the duct in another and abated with supportive care in the other.