The Effects of Vehicle Movement/ Operation on Fluid Flow Regulators During Patient Transport.
- Author:
Hyun Jae PARK
1
;
Jung Weon LEE
;
Sae Hoonn PARK
;
Jae Woo KIM
;
Young Soon CHO
Author Information
1. Department of Emergency Medicine, College of Medicine, Soonchunhyang University, Cheonan, Korea. mdgarden@gmail.com
- Publication Type:Original Article
- Keywords:
Infusion pump;
Transportation of patients;
Fluid therapy
- MeSH:
Fluid Therapy;
Humans;
Infusion Pumps;
Transportation of Patients
- From:Journal of the Korean Society of Emergency Medicine
2012;23(1):74-77
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: Evaluation of the effects of a vehicle's movement/ operation on fluid flow regulators during patient transport. To determine whether or not vehicle trembling during idling and movement during patient transport are factors affecting the velocity of fluid injection. METHODS: The volume of fluid, as measured in an idling or moving vehicle, was evaluated using three different types of marketed flow regulators at three different injection speeds: 10, 20 and 40 ml/hr. RESULTS: In all cases, when the vehicle was idling or in motion, discrepancies were observed between the pre-calculated amount of fluid and the actual amount of fluid injected. However, a greater discrepancy was observed to exist in a moving vehicle. CONCLUSION: The trembling and movement of a vehicle during patient transport affects fluid injection results.