A study on fitness of several domestic implant fixture and abutment screws.
- Author:
Young Sun PARK
1
;
Mong Sook VANG
;
Seok Hyung LEE
Author Information
1. Department of Prosthodontics, College of Dentistry, Chonnam National University.
- Publication Type:Original Article
- Keywords:
Domestic implants;
Fit;
Screw;
Scanning electron microscope
- MeSH:
Dental Implants
- From:The Journal of Korean Academy of Prosthodontics
2004;42(1):94-109
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
STATEMENT OF PROBLEM: Several domestic dental implants have been developed since dental treatment being implants are increasing and popularized. However, they have not been used in domestic market like imported goods. MATERIAL AND METHODS: This study was carried out to compare domestic and imported implants in aspect of fitness, dental implant fixture and abutment screw using scanning electron microscope. All experiments were performed under dry condition. RESULTS: 1. Only in aspect of relation of dental implant fixture and abutment screw, except only one group with point contact, good fitness was existed. 2. Home products must presevere in their efforts, so as excellent to fit.