Normal Value of Standard Electroretinography and Change with Age and Sex(II)-Results Using Burian-Allen Electrode and Comparison with Results Using EFG-jet Electrode-.
- Author:
Shung Hee CHOI
;
Young Hoon OHN
;
Hanho SHIN
- Publication Type:Original Article
- Keywords:
Burian-Allen electrode;
Electroretinogram
- MeSH:
Electrodes*;
Electroretinography*;
Female;
Humans;
Reference Values*;
Retinal Diseases
- From:Journal of the Korean Ophthalmological Society
1999;40(1):128-139
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
Using Burian-Allen bipolar contact lens electrode, Authors examined the normal 120 eyes of 120 subjects(male 57, female 63), age ranged 7 to 83 years(mean 44.5 years), with the method which was same as the ISCEV standard. There was no statistically significant differences between the value obtained using Burian-Allen electrode and the value obtained using ERG-jet electrode(p>0.05). We expect that this data would be useful testing method to evaluate the retinal disease. And this data may provide new reference value in the study of electroretinogram.