Investigation of typical values in cardiovascular interventional diagnosis and treatment and assessment of radiation protection optimization: a single-center cross-sectional study
10.13491/j.issn.1004-714X.2025.04.010
- VernacularTitle:心血管介入诊断与治疗典型值调查与放射防护优化评估:一项单中心横断面调查
- Author:
Guoliang JIN
1
;
Limeng CUI
2
;
Xiaolong MA
1
;
Xiaohai MA
1
;
Zechen FENG
2
;
Yongzhong MA
2
Author Information
1. Beijing Anzhen Hospital, Affiliated to Capital University of Medical Sciences, Beijing 100029, China.
2. Beijing Center for Disease Prevention and Control, Beijing 100013, China.
- Publication Type:OriginalArticles
- Keywords:
Diagnostic reference level;
Typical value in institution;
Interventional radiology;
Radiation protection
- From:
Chinese Journal of Radiological Health
2025;34(4):530-533
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish typical values for interventional diagnosis and treatment at our institution, use these values as a tool to evaluate patient medical exposure doses, and optimize radiation protection measures. Methods From June to December 2023, we collected information on 593 adult cardiovascular interventional diagnosis and treatment surgeries, including surgery type, equipment model, air kerma-area product (KAP), incident reference point air kerma (Ka,r), perspective time (FT), and exposure mode. Results The typical value of cardiovascular interventional diagnosis at our institution in 2023 was 27.5 Gy·cm2. The typical value of cardiovascular interventional treatment was 70.0 Gy·cm2. The FT, KAP, and Ka,r of interventional surgeries were significantly higher than those of interventional diagnosis (P < 0.01). There were significant correlations between FT, KAP, and Ka,r (P < 0.01). Conclusion The results of this study were slightly different from those of other studies. They provide typical data and reference values for cardiovascular interventional diagnosis and treatment dose levels in Beijing and are helpful for dose optimization between different medical institutions.