Delayed Aneurysm Rupture Following Endovascular Treatment with Contour Device: A Case Report
10.5469euroint.2024.00493
- Author:
Osman KOC
1
;
Mostafa MAHMOUD
;
Ehab MAHMOUD
;
Ali AYYAD
;
Ahmad OWN
Author Information
1. Department of Neuroradiology, Neuroscience Institute, Hamad Medical Corporation, Hamad General Hospital, Doha, Qatar
- Publication Type:Case Report
- From:Neurointervention
2025;20(1):24-27
- CountryRepublic of Korea
- Language:English
-
Abstract:
Delayed rupture of intracranial aneurysms after endovascular treatment is a rare but serious complication. We report the first documented case of late aneurysmal rupture following treatment with a Contour intrasaccular device. A patient in their 60s with a basilar tip aneurysm underwent endovascular treatment using a 14-mm Contour device. Fifteen months later, the patient presented with a fatal intraventricular hemorrhage, and imaging revealed device displacement and aneurysm growth. This case underscores the importance of meticulous device sizing and follow-up, especially for large aneurysms.