Evaluation of application of enzyme immunoassay workstation in detection of antibodies against Streptococcus pneumoniae
10.13200/j.cnki.cjb.004430
- VernacularTitle:酶免工作站在肺炎链球菌抗体检测中的应用评估
- Author:
ZHOU Shanshan
- Publication Type:Journal Article
- Keywords:
Enzyme immunoassay workstation;
Streptococcus pneumoniae(S.pneumoniae);
IgG antibodies
- From:
Chinese Journal of Biologicals
2025;28(02):210-214+220
- CountryChina
- Language:Chinese
-
Abstract:
Objective To apply automated enzyme immunoassay workstation instead of manual operation to facilitate detection of Streptococcus pneumoniae(S.pneumoniae) IgG antibodies, so as to improve the detection efficiency.Methods Procedures of the assay were set up on the automated workstation according to the World Health Organization(WHO) Pneumococcal Capsular Polysaccharide Types-pecific IgG Antibody Detection Manual Pn PS ELISA. The number of working station panels was maximized to preset all reagent consumables and more samples to be tested, thereby improving the detection flux and reducing the number of manual interventions. The serum PnG was repeatedly detected to verify the test precision. The WHO S.pneumoniae international calibration serum panel 12/278 was detected to verify the accuracy of the test. To compare the consistency of workstations and manual operation, 16 same serum samples were tested by manual operation and on the workstation meanwhile. Detections of four calibration serum samples were performed on the workstation by double-(WHO standard method) and single-column methods, and the consistency of the results was compared.Results With the layout of 21 plate positions, 24 serum samples could be examined in one program run, requiring only 2 manual interventions. The CV of 32 detection results of serum PnG was lower than 15%. The assay results of 75% of the calibration serum panel in 14 out of 23 serotypes exhibited a percent error of 40% or less compared to the assigned values, and the total accuracy was 80%.The correlation coefficient(Pc) of agreement between workstation and manual test results was 0. 996 7, and the correlation coefficient of agreement between workstation single-and double-column test results was 0. 999 8, both of which showed high agreement.Conclusion The automated enzyme immunoassay workstation has shown high precision, accuracy and large detection throughput, which is labor-saving, and can be applied to the detection of S.pneumoniae antibodies. The results of single-and double-column assay provide experimental basis for single-column detection in place of double-column detection.