Liquid chromatography-tandem mass spectrometry (LC-MS/MS) based assay for the quantification of melatonin in human serum
10.3760/cma.j.cn114452-20231107-00264
- VernacularTitle:液相色谱-串联质谱检测血清褪黑素方法建立和临床应用研究
- Author:
Yingfei PENG
1
;
Zhenxin WANG
;
Fangjun CHEN
;
Jiaqian QIN
;
Baishen PAN
;
Beili WANG
;
Wei GUO
Author Information
1. 复旦大学附属中山医院检验科,上海 200032
- Keywords:
Mass spectrometry;
Liquid chromatography-tandem mass spectrometry;
Melatonin;
Performance verification
- From:
Chinese Journal of Laboratory Medicine
2024;47(7):812-818
- CountryChina
- Language:Chinese
-
Abstract:
Objective:The aim of this study was to develop and validate a new liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for the quantification of melatonin in human serum.Methods:We describe the performance and validation of melatonin by LC-MS/MS. 182 serum samples from the patients diagnosed with Sleep disturbance who visited the Department of Psychiatry at Zhongshan Hospital affiliated to Fudan University from February 2022 to March 2023(56 males,162 females,mean age [45.51±16.31]years), as well as 182 healthy individuals were included(87 males,95 females,mean age [48.55±11.93]years). The two groups were used to assess the application of serum melatonin levels as a diagnostic indicator for sleep disorders (SDs). The liquid chromatography mass spectrometry (LC-MS) system with an chromatography column (2.1×100 mm, 1.8 μm) was used for separation. The column temperature was set at 35 ℃, as well as the mobile phase consisting of a 0.1% formic acid aqueous solution and pure acetonitrile. The flowing rate was set at 0.4 ml/min for gradient elution. The LC-MS/MS method was validated according to guidance documents, including the following parameters: specificity, selectivity, matrix effect, carryover contamination and reproducibility, lower limit of measuring interval, linearity, precision, recovery rate, dilution consistency, and serum sample stability. Then, it was subsequently employed to profile melatonin changes in Sleep disturbance.Results:The lower limit of quantification for melatonin was 1 pg/ml, and the linear range of detection was 1 pg/ml to 500 pg/ml ( r=0.999). The intra-day and intra-batch precision, expressed as the coefficient of variation ( CV), was within the range from 3.07% to 6.86%, which met the requirement of less than 15%. The recovery rate of the spiked samples ranged from 105.91% to 116.30%. The level of serum melatonin in the sleep disturbance group was significantly lower than that in the healthy control group ([2.00(1.00,3.28)] vs [8.35(4.28,14.80)] pg/ml, P<0.001). Conclusions:The LC-MS/MS method we developed for the quantification of melatonin is clinical practicable.