Phantom study based on MRI cine sequences: analysis of the accuracy of tumor motion range accuracy
10.3760/cma.j.cn113030-20230720-00013
- VernacularTitle:基于MRI电影序列的模体研究:肿瘤运动范围准确性分析
- Author:
Bing LI
1
;
Yuan WANG
;
Ronghu MAO
;
Dong LIU
;
Wenzheng SUN
;
Xiangyue LIU
;
Nan MENG
;
Wei GUO
;
Shuangliang CAO
;
Xipan LI
;
Chen CHENG
;
Hui WU
;
Hongyan TAO
;
Dingjie LI
;
Zhaoyang LOU
;
Hongchang LEI
;
Lingguang MENG
;
Hong GE
Author Information
1. 河南省肿瘤医院放疗科,郑州大学附属肿瘤医院,中国医学科学院肿瘤医院河南医院,郑州 450008
- Keywords:
Radiotherapy;
Magnetic resonance cine sequences;
Tumor motion;
TrueFISP
- From:
Chinese Journal of Radiation Oncology
2024;33(12):1144-1151
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the accuracy of magnetic resonance imaging (MRI) cine sequences in determining the range of tumor motion in radiotherapy, providing a basis for the precise delineation of the target volume in motion for radiation therapy.Methods:A modified chest motion phantom was placed in a MRI scanner, and a water-filled sphere was used to simulate a tumor. True fast imaging with steady precession (TrueFISP) MRI cine sequences from Siemens were used to capture the two-dimensional motion images of the simulated tumor. The phantom experiments were divided into three modes: head-foot motion mode, rotation motion mode, and actual respiratory waveform mode. In the head-foot motion mode, respiratory motion period (3, 4, 5, 6, 7 and 8 s), amplitude (5, 10 and 15 mm), and respiratory waveform of the simulated tumor (sin and cos4) were set, resulting in a total of 36 motion combinations. In the rotation motion mode, a cos4 waveform was used for respiration, with respiratory periods of 3, 4, 5, 6, 7 and 8 s, head-foot motion set amplitudes of 5, 10 and 15 mm, and anterior-posterior (AP) and left-right (LR) motion set amplitudes in three combinations ([2.5, 2.5] mm, [2.5, 5.0] mm, [5.0, 5.0] mm), resulting in a total of 54 motion combinations. In the actual respiratory waveform mode, respiratory waveforms of 5 randomly selected patients from Affiliated Cancer Hospital of Zhengzhou University were obtained. Under each motion combination, TrueFISP cine images (30 frames, with an acquisition time of 11 s per frame) were obtained. The code was used to automatically identify the two-dimensional coordinates of the center of the simulated tumor in each image, and sin and cos4 functions were separately employed to fit the tumor position in the motion direction, thereby obtaining the fitted motion period and amplitude. The difference between the maximum and minimum values of the tumor's center coordinates in the head-to-foot direction is taken as the range of movement, referred to as the calculated amplitude. For the actual respiratory waveform, the distance between the measured maximum and minimum positions is used to calculate the amplitude.Results:In the head-foot motion mode, the fitted amplitudes of both sin and cos4 waveforms deviated from the set amplitudes by 0-0.51 mm, with relative deviations of 0%-4.2%. The deviation range between the calculated amplitudes and the set amplitudes of the two waveforms were 0.08-0.94 mm, with relative deviations of 1.1%-6.3%. In the rotation motion mode, the fitted amplitudes deviated from the set amplitudes by 0-0.61 mm, with relative deviations of 0%-6.2%. And the deviation range between the calculated amplitudes and the set amplitudes were 0.16-0.94 mm, with relative deviations of 0%-6.3%. In the actual respiratory waveform motion mode, the deviation range between the calculated amplitudes and the set amplitudes were 0.10-0.48 mm, with relative deviations of 2.2%-8.6%.Conclusion:TrueFISP cine sequences show minimal deviations in determining the range of tumor head-foot motion and effectively captures the tumor's movement state, thereby providing important support for the precise definition of the tumor movement target area during radiotherapy .