Study on the distribution characteristics of radiation dose of body surface of operators in cardiovascular interventional treatment
10.3969/j.issn.1672-8270.2024.08.004
- VernacularTitle:心血管介入手术医师体表辐射剂量分布特征研究
- Author:
Yan ZHANG
1
,
2
;
Haibo HUANG
;
Yiwei SU
;
Peng RAO
;
Shifeng HOU
;
Zhi WANG
;
Jianyu WANG
Author Information
1. 广州市第十二人民医院职业环境与健康重点实验室 广州 510620
2. 广州医科大学公共卫生学院 广州 510000
- Keywords:
Operator of interventional operation;
Radiation dose of body surface;
Characteristics of dose distribution;
Radiation protection
- From:
China Medical Equipment
2024;21(8):18-22
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To analyze and verify the distribution characteristics of radiation dose on the body surface of surgeons in cardiovascular interventional treatment,so as to provide theoretical basis for studying the characteristics of radiation dose and implementing radiation protection correctly.Methods:During cardiovascular interventional surgery,three pieces of lead protective clothing(designated as lead protective clothing A,B,and C)were provided in the room of digital subtraction angiography(DSA)equipment.Each piece of these protective clothing included one lead scarf,one lead vest and one lead apron,and the lead equivalents of them were all 0.5 mmPb.A thermoluminescence dosimeter was placed on the outer surface of the lead protective clothing that was worn by surgeon,and the radiation dose values of the front and back,vertical and horizontal surfaces of the three pieces of lead protective clothing were compared.Results:The measured average values of the front dosimeter of the three pieces of lead protective clothing were respectively(296.2±31.9)mSv,(196.3±20.2)mSv and(299.4±33.0)mSv,which were significantly higher than those of the back of the lead protective clothing,and the differences were statistically significant(Z=-5.522,-5.115,-4.817,Z=-5.522,-5.115,-4.817,P<0.05).The values of abdominal radiation doses on the vertical plane of the three pieces of lead protective clothing were the highest,which were respectively(356.7±58.2)mSv,(217.7±31.8)mSv and(387.4±46.3)mSv.The measured values of the dosimeter on the left side of the horizontal plane of the A,B of lead protective clothing were respectively(341.6±58.5)mSv and(229.1±32.4)mSv,which were all higher than that on the right side,with statistical significances(Z=5.890,9.068,P<0.05).The measured value of the left side of lead protective clothing was respectively higher than that of the middle side and the right side,and the differences were statistically significant(Z=5.846,9.766,P<0.05).Conclusion:The radiation dose of the front of the body surface of surgeon who undergo DSA intervention is significantly higher than that in the back of that,and the radiation dose of the abdomen on the vertical plane of the body is the highest,and the radiation dose of horizontal plane appears a decreasing trend from left to right.The detected dose data and distribution characteristics of radiation can provide research basis for radiation protection.