Study on the individualized dose verification of patients with CyberKnife treatment based on dose verification system of SRS MapCHECK matrix
10.3969/j.issn.1672-8270.2024.07.002
- VernacularTitle:基于二维半导体矩阵剂量验证系统实施射波刀治疗患者个体化剂量验证研究
- Author:
Hanshun GONG
1
;
Shanshan GU
;
Shaojuan WU
;
Jinglin SUN
;
Pengfei XU
;
Xiaoliang LIU
;
Jingmin BAI
;
Chuanbin XIE
Author Information
1. 解放军总医院第一医学中心放射治疗科 北京 100853
- Keywords:
CyberKnife(CK);
Two-dimensional semiconductor matrix(SRS MapCHECK);
Quality assurance(QA);
γ passing rate
- From:
China Medical Equipment
2024;21(7):17-22
- CountryChina
- Language:Chinese
-
Abstract:
Objective:A dose verification system of two-dimensional semiconductor matrix(SRS MapCHECK)was used to verify the dose of the clinical treatment plan of patients who underwent CyberKnife(CK),which realized rapid verification for individualization of radiotherapy plans of patients through analyzed the γ-passing rates of them.Methods:A total of 253 patients with tumor who received CK clinical treatment in the First Medical Center of Chinese PLA General Hospital from March 2021 to May 2023 were selected.Among of them,121 cases received CK treatment on head,and 30 cases received that on lung,and 102 cases received CK treatment on abdomen and other metastatic tumor.In the MultiPlan treatment plan system,the plan of patient was mapped to the integrated model composed of StereoPHAN model and SRS MapCHECK matrix dose verification system by the means of the plan image center overlap.The dose verification was conducted on the plan of each patient on the basis of ensuring the consistency of the number of beam,direction of beam and the monitor unit.The different γ analysis standards(1%/1 mm,2%/1 mm,3%/1 mm,1%/2 mm,2%/2 mm,3%/2 mm,1%/3 mm,2%/3 mm and 3%/3 mm)were adopted to conduct global analysis of absolute dose for each verification plan,and the threshold(TH)of low dose was set as 10%.Results:The γ passing rates of phantom verification plans of 253 patients were respectively(88.64±5.91)%,(95.43±3.40)%,(97.90±2.06)%,(96.51±2.35)%,(98.15±1.68)%,(99.06±1.12)%,(98.30±1.39)%,(99.09±0.97)%and(99.52±0.63)%under different analysis standards.The γ passing rates of other standards of patients with tumor on different parts were larger than 95%except the analysis result of 1%1 mm standard.The overall analysis result of the deviation of central point dose was(-1.30±2.17)%,among of which the tumor of head,abdominal tumors and other metastatic tumor were about approximately-2%,while that of lung tumors were approximately-3%.The deviation of abdominal and other metastatic tumor was the minimum.The correlation analysis showed that the target volume and the size of the minimum collimator were respectively correlated to the dose deviation of the center.Conclusion:SRS MapCHECK dose verification system can conveniently and quickly realize the individualized verification for the plan of patients who receive CK treatment.