Comparative study of POCT instruments and Roche electrochemiluminescence instrument in detecting myocardial biomarkers
10.3969/j.issn.1672-8270.2024.05.003
- VernacularTitle:即时检测仪器与电化学发光仪检测心脏标志物的对比研究
- Author:
Wenjuan LIU
1
;
Pan GUO
;
Juan DU
;
Herui ZHANG
;
Yaping WANG
;
Hailong WANG
;
Jia XU
;
Qi LI
Author Information
1. 中国中医科学院西苑医院检验科 北京 100091
- Keywords:
Point of care testing(POCT);
Myocardial biomarkers;
Electrochemiluminescence;
Analysis performance
- From:
China Medical Equipment
2024;21(5):16-19,25
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To compare and analyze the correlation between two kinds of point-of-care testing (POCT) instruments and Roche electrochemiluminescence instrument in detecting myocardial biomarkers,so as to provide reference for the selection of POCT instruments in clinical testing. Methods:A total of twenty patient's samples were selected,which concentration range covered the reportable range of detection results. Two kinds of POCT instruments (named respectively A and B) were used to detect the concentrations of plasma myoglobin (Myo),creatine kinase isoenzyme mass (CK-MB mass),cardiac troponin I(cTnI),and N-terminal fragment of B-type natriuretic peptide (NT-proBNP)/Brain natriuretic peptide (BNP). The results of the electrochemiluminescence instrument were used as reference,and Pearson correlation analysis and Kappa test evaluation were used to evaluate respectively the correlation and consistency of the detected results between the two kinds of POCT instruments and Roche electrochemiluminescence instrument. Results:Our study demonstrated that there were good correlations in the detected results of Myo,CK-MB mass,cTnI,NT-proBNP/BNP between the instrument A and Roche electrochemiluminescence instrument,and between the instrument B and Roche electrochemiluminescence instrument (rinstrumentA=0.994,0.989,0.917,0.996,rinstrumentB=0.928,0.934,0.883,0.977,P<0.05),respectively. In the comparisons about consistencies,the concordance rates between instrument A and electrochemiluminescence instrument were respectively 90%,100%,70% and 100% in detecting Myo,CK-MB,cTnI and NT-proBNP,and the concordance rates between instrument B and electrochemiluminescence instrument were respectively 90%,100%,70% and 100% in detecting them. Both two kinds of POCT instruments had favorable consistencies with Roche electrochemiluminescence instrument (Kappainstrument A=0.798,1.00,0.429,1.00,Kappainstrument B=0.794,0.886,0.429,1.00),respectively. The precisions of both instruments were less than 10%. Conclusion:Both two kinds of POCT instruments can meet the requirements of clinical test,which have favorable correlations with electrochemiluminescence instrument in detecting myocardial biomarkers.