5.0T MR for cardiac imaging:Comparison with 3.0T MR
10.13929/j.issn.1003-3289.2024.05.006
- VernacularTitle:5.0T MR心脏成像:与3.0T MR对比
- Author:
Lan LAN
1
;
Naili YE
;
Huijuan HU
;
Wenbo SUN
;
Rongqing SUN
;
Gonghao LING
;
Tingyi DU
;
Xuan LI
;
Xiaopeng SONG
;
Haibo XU
Author Information
1. 武汉大学中南医院医学影像科,湖北武汉 430071
- Keywords:
heart;
magnetic resonance imaging;
prospective studies
- From:
Chinese Journal of Medical Imaging Technology
2024;40(5):661-665
- CountryChina
- Language:Chinese
-
Abstract:
Objective To observe the feasibility of 5.0T MR for cardiac imaging.Methods Three patients with heart diseases and 17 healthy volunteers were prospectively enrolled.Cardiac MR(CMR)cine sequence and black blood sequence imaging were performed using 5.0T and 3.0T MR scanner,respectively.The image quality and artifacts degrees were compared between 5.0T and 3.0T CMR images,and the consistency of left ventricular parameters obtained using 5.0T and 3.0T scanners was analyzed.Results No significant difference of image quality nor artifacts degrees was found between 5.0T and 3.0T CMR images(all P>0.05).The left ventricular end diastolic volume(EDV),end systolic volume(ESV),ejection fraction(EF),stroke volume(SV)and end diastolic mass(EDM)derived from cine images acquired at different fields were in a good agreement(all ICC>0.75,all P<0.001).Conclusion 5.0T MR could be used for cardiac imaging,with image quality of cine and black blood sequences comparable to that of 3.0T MR.