Factors affecting MV imager projection offset in machine performance check for Varian linear accelerator
10.3969/j.issn.1005-202X.2024.07.004
- VernacularTitle:瓦里安直线加速器机器性能检查中MV影像中心偏移量影响因素分析
- Author:
Liuyang XU
1
;
Xiaoyin WANG
;
Shouyu WANG
;
Kehua PANG
;
Dandan SUN
;
Jun YANG
Author Information
1. 新乡医学院第一附属医院放射治疗技术科,河南卫辉 453100
- Keywords:
linear accelerator;
machine performance check;
on-board imager;
encoder;
isocenter calibration
- From:
Chinese Journal of Medical Physics
2024;41(7):808-812
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the main factors affecting the MV imager projection offset in the machine performance check(MPC)for Varian Vital Beam linear accelerator.Methods The MV imager projection offsets in the MPC after repairing the MV imaging arm encoder of shoulder motor,locking the treatment couch,and isocenter calibration were analyzed.Results MPC results revealed that the MV imager projection offset after repairing the MV imaging arm encoder of shoulder motor was(0.310±0.001)mm,significantly less than(0.450±0.010)mm in the blank group.The difference in MV imager projection offset between the isocenter calibration group and the blank group was trivial.The MV imager projection offset after locking the treatment couch was(0.240±0.030)mm,significantly less than(0.450±0.010)mm in the blank group.When MPC was carried out after repairing the imaging arm encoder and performing isocenter calibration,there was no significant statistical difference in MV imager center offset between the locked and unlocked treatment couch.Conclusion The damage of MV imaging arm encoder of shoulder motor is the main factor causing abnormal MV imager projection offsets.Locking the treatment couch before the MV imaging center check can reduce the results,but it cannot eliminate the MV imager projection offset.