Optimization of simulated localizable CT scanning parameters for pediatric body radiotherapy
10.3969/j.issn.1005-202X.2024.05.002
- VernacularTitle:儿童体部放疗模拟定位CT扫描参数优化
- Author:
Guiyuan LI
1
;
Xinyuan CHEN
;
Liang ZHAO
;
Xiaohong WANG
;
Yuan TIAN
;
Jianrong DAI
;
Junlin YI
Author Information
1. 国家癌症中心/国家肿瘤临床医学研究中心/中国医学科学院北京协和医学院肿瘤医院廊坊院区,河北廊坊 065001
- Keywords:
simulated localizable CT;
children;
radiotherapy;
scanning parameter;
parameter optimization
- From:
Chinese Journal of Medical Physics
2024;41(5):534-540
- CountryChina
- Language:Chinese
-
Abstract:
Objective On the premise of meeting the image quality requirements of simulated location for pediatric radiotherapy,the simulated localizable CT parameters are optimized through phantom scanning to reduce the radiation dose.Methods CatPhan700 phantom was used to simulate the child's body,Philip 24-row large-aperture spiral simulated localizable CT was performed,and the CT images were obtained by scanning the phantom at different mAs and tube voltages.The mAs range was set at 60-400 mAs,the scanning was performed every 20 mAs interval,and the kV was set at 80,100,and 120 kV.Image evaluation was carried out using parameters such as image noise(N10 and mean SD),uniformity,low contrast resolution,high contrast resolution,and the stabilities of HU values of Air,Acrylic,50%bone,LDPE,20%bone,Teflon,Polystyrene,DelrinTM,Lung,PMP and Water.The CTDIVol and DLP automatically calculated by the simulated localizable CT system were read to evaluate the radiation dose.Results At 100 kV,as mAs increased,both CTDI and DLP showed upward trends,and the fitting results were linear correlated,with slopes of 0.034 5 and 0.932 4.Image noise was decreased nonlinearly with the increasing mAs.When mAs increased from 60 to 140 mAs,N10 decreased from 0.25%to 0.14%,and SD reduced from 3.74 HU to 2.54 HU.When mAs reached 180 mAs or higher,N10 fluctuated between 0.1%and 0.12%,the mean SD fluctuated between 2.0 and 2.5 HU,and the downward trends obviously slowed down.When mAs increased from 60 to 200 mAs,the low contrast resolution of the image dropped from 0.53 to 0.29.The image uniformity,high contrast resolution and HU values of different substances were less affected by mAs.The image quality of 100 kV and 200 mAs scanning was close to that of 120 kV scanning,but the image quality of 80 kV scanning failed to meet the clinical requirements.Conclusion In order to reduce the radiation dose as much as possible,the mAs should be set at 200 mAs when the tube voltage is set at 100 kV for a simulated cylinder with a diameter of 20 cm.In the actual simulation scanning for pediatric radiotherapy,the scanning parameters should be fine-tuned according to the phantom results and the actual physical characteristics of children to satisfy the optimization principle for radiation protection.