Determination of aluminum content in hemofiltration base solution by graphite furnace atomic absorption standard addition method
10.19778/j.chp.2024.02.015
- VernacularTitle:石墨炉原子吸收标准加入法测定血液滤过置换基础液中的铝含量
- Author:
Zhen LI
1
;
Weijie YU
;
Zenghui LUAN
;
Yunjie ZHU
;
Zhao YANG
Author Information
1. 青岛市食品药品检验研究院,山东青岛 266000
- Keywords:
graphite furnace atomic absorption method;
standard addition method;
hemofiltration basic solution;
aluminum;
pollution;
limit control
- From:
Drug Standards of China
2024;25(2):200-203
- CountryChina
- Language:Chinese
-
Abstract:
Abtract Objective:To establish a graphite furnace atomic absorption spectroscopy,and to determine the content of aluminum in the hemofiltration base solution.Methods:The standard addition method of atomic absorption of graphite furnace was used to add matrix modifier to determine the content of aluminum.Results:Aluminum has a good linear relationship in the range of 0-20 μg·L-1,r=0.998;The detection limit concentration was 0.91 μg·L-1.The average recovery rate was 96.5%.The results of the three batches were 3.299,1.232 and 2.431 μg·L-1,respectively.Conclution:This method can effectively measure the content of aluminum in hemofiltration base solution products,and control the raw materials,production and packaging of products that may introduce pollution pathways.It is recommended that enterprises pay attention to the detection of aluminum content in the on hemofil-tration base solution to minimize the risk of contamination and ensure the quality of products.