Comparison of Long-Term Results between One-piece and Three-piece Acrylate Intraocular Lens.
10.3341/jkos.2008.49.2.245
- Author:
Hwan HEO
1
;
Hwang Gyun KIM
;
Kyung Chul YOON
;
Yeoung Geol PARK
Author Information
1. Department of Ophthalmology, Chonnam National University Medical School & Hospital, Chonnam National Research Institute for Medical Sciences, Gwangju, Korea. kcyoon@chonnam.ac.kr
- Publication Type:Original Article
- Keywords:
Axial displacement;
Hydrophobic acrylic intraocular lens;
Refractive stability
- MeSH:
Acrylates;
Anterior Chamber;
Astigmatism;
Capsule Opacification;
Displacement (Psychology);
Endothelial Cells;
Eye;
Follow-Up Studies;
Humans;
Incidence;
Intraocular Pressure;
Iris;
Lenses, Intraocular;
Retrospective Studies;
Visual Acuity
- From:Journal of the Korean Ophthalmological Society
2008;49(2):245-252
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: To compare the long-term clinical results of one-piece Acrysof(R) (SA60AT) hydrophobic acrylic intraocular lens (IOL) implantation compared with implantation of three-piece Acrysof(R) (MA60BM) hydrophobic acrylic IOL. METHODS: We retrospectively analyzed each 50 eyes of 50 patients underwent MA60BM or SA60AT IOL implantation and followed for at least 6 months. RESULTS: Final visual acuity of 0.5 or better was 38 eyes (76%) and 0.8 or better was 20 eyes (40%) in the SA60AT group. In the MA60BM group, it was 41 eyes (82%) and 23 eyes (46%) respectively. There were no significantly differences in predictability, intraocular pressure, endothelial cell density, astigmatism, and incidence of posterior capsule opacification between the two groups. Spherical equivalent at postoperative 1 week was -0.50+/-0.95D in SA60AT group and was -0.31+/-0.88D in MA60BM group (P=0.04). However, there was no significant difference between the two groups during follow up period. In MA60BM group, anterior chamber depth (P=0.02) and distance between iris and IOL (P=0.04) reduced significantly during the first postoperative month. CONCLUSIONS: Early postoperative axial displacement and changes in spherical equivalent can be occurred in MA60BM group. However there was no difference in long-term clinical results between SA60AT and MA60BM groups.