Application of remote sensing navigation system in total knee arthroplasty
10.12200/j.issn.1003-0034.20240342
- VernacularTitle:遥感导航系统在全膝关节置换术的应用
- Author:
Yuan-Yuan LI
1
;
Ming-Jiang HE
;
Peng-Cheng SHAN
;
Pei-Yan HU
;
Lin JING
;
Qi YAN
;
Hai TANG
;
Xin-Yue WANG
;
Si-Ye LIU
;
Hong-Mei ZHANG
Author Information
1. 中国中医科学院望京医院骨关节一科,北京 100102
- Keywords:
Total knee arthroplasty;
Remote sensing navigation alignment;
Prosthesis alignment;
Fiducial markers
- From:
China Journal of Orthopaedics and Traumatology
2024;37(9):878-885
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore clinical accuracy of remote sensing navigation alignment(RSNA)system in total knee arthroplasty(TKA)and its influence on postoperative clinical efficacy.Methods From May 2021 to May 2022,60 knee os-teoarthritis(KOA)patients with Kellgren-Lawrence(K-L)grade Ⅲ to Ⅳ treated by unilateral primary TKA were selected and divided into RSNA group and traditional operation group according to treatment methods,and 30 patients in each group.There were 6 males and 24 females in RSNA group,aged from 55 to 86 years old with an average of(68.06±8.23)years old;body mass index(BMI)ranged from 22.15 to 34.58 kg·m-2 with an average of(28.20±3.01)kg·m-2;the courses of disease ranged from 2 to 60 months with an average of(18.80±14.80)months;13 patients with grade Ⅲ and 17 patients with grade Ⅳaccording to K-L grading.In traditional operation group,there were 8 males and 22 females,aged from 57 to 85 years old with an average of(67.26±6.32)years old;BMI ranged from 23.94 to 34.55 kg·m-2 with an average of(27.49±2.32)kg·m-2;the courses of disease ranged from 3 to 60 months with an average of(21.30±16.44)months;14 patients with grade Ⅲ and 16 pa-tients with grade Ⅳ according to K-L grading.Western Ontario and McMaster Universities(WOMAC)osteoarthritis index and Knee Society score(KSS)were used to evaluate functional recovery of patients.Hip-knee-ankle angle(HKAA),distal femoral valgus angle(FVA)and distal fermoral flexion angle(DFFA)were measured before operation.HKAA and HKAA deviation angle were measured at 1 week after operation,and defective rate of lower limb force line,femur prosthesis valgus angle(FP-VA)and femoral prosthesis flexion angle(FPFA),respectively,were calculated.Results There were no serious complications such as vascular and nerve injury during operation,and wound healed at stage Ⅰ.Both groups were followed up for 6 months.There were no significant difference in WOMAC index,KSS,HKAA,FVA and DFFA between two groups before operation(P>0.05).The force line defect rate,HKAA,HKAA deviation angle,FPVA deviation angle and FPFA of RSNA group were 6.7%,(178.74±1.56)°,(1.25±1.56)°,(1.84±0.16)° and(4.85±2.46)°,respectively;while in traditional operation group were 20%,(176.73±3.46)°,(3.27±3.46)°,(2.44±0.26)°,(6.60±1.86)°;the difference between two groups were statistically sig-nificant(P<0.05).There were no significant difference in WOMAC index and KSS between two groups at 3 and 6 months after operation(P>0.05).Conclusion RSNA system could reduce defective rate of lower limb force line,FPVA deviation angle and FPFA after TKA,which is more accurate and easy to operate than traditional intramedullary localization surgery while ensuring postoperative efficacy.