Development of low-oxygen mixed gas generator for pilot hypoxia testing
10.19745/j.1003-8868.2024126
- VernacularTitle:飞行员缺氧耐力检查用低氧混合气体制备装置的研制
- Author:
Lin-Xia LI
1
;
Jia-Ling XU
;
Guo-Yun MAO
;
Yao-Xuan JI
;
Jin MA
;
Yun-Ying WANG
Author Information
1. 空军军医大学航空航天医学系航空航天生理学教研室,西安 710032
- Keywords:
hypoxia tolerance testing;
low-oxygen gas preparation;
acclimatization training;
flight safety
- From:
Chinese Medical Equipment Journal
2024;45(7):24-28
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a low-oxygen mixed gas generator to make up for the deficiencies of low-pressure chambers and load-resistant hypoxia trainers during pilot hypoxia tolerance testing.Methods The device prepared the low-oxygen gas with the principle of gas separation,which was composed of a Sunsource OLF1100D-220AF air compressor,a SMC IDG75SAM4-03 filter,a buffer tank,an AIR Products PA3010 integrated assembly,a control box,sensors and regulators.The sensors included the pressure sensor,flow sensor,concentration sensor and dew point temperature sensor,and the control box consisted of a main control board,a power supply module,a transmission module,a communication module,a display and a housing.The embedded control software of the device was developed with KEIL 5 and C++.Results The device developed prepared the low-oxygen gas with the volume fraction being 4%to 18%and the maximum error of volume fraction being 0.05%,and the main components of the prepared gas met the technical requirements of medical oxygen as stipulated in GB 8982-2009 Oxygen supplies for medicine and aircraft breathing.Conclusion The low-oxygen gas prepared by the device has its volume fraction precisely controlled and can be used for hypoxia tolerance testing and acclimation training for pilots.[Chinese Medical Equipment Journal,2024,45(7):24-28]