Study on fatigue vibration evaluation of ultrasonic knife tip based on Q factor
10.19745/j.1003-8868.2024106
- VernacularTitle:基于品质因数的超声刀刀头疲劳振动评价研究
- Author:
Ke-Sheng WANG
1
;
Ze-Kai LI
;
Pei LIU
;
Jing-Sheng SUN
;
Xu-Guang PENG
;
Shuang-Shuang LI
;
Qian-Hong HE
;
Zhen LIU
Author Information
1. 电子科技大学-水木医疗联合实验室,成都 611731
- Keywords:
ultrasonic knife;
Q factor;
fatigue experiment;
vibration fatigue;
laser vibrometer;
non-contact measurement
- From:
Chinese Medical Equipment Journal
2024;45(6):17-22
- CountryChina
- Language:Chinese
-
Abstract:
Objective To propose a Q factor-based fatigue vibration evaluation method of the ultrasonic knife tip.Methods Firstly,an ultrasonic cutter fatigue testing table was established to realize repeated cutting,which was composed of a power supply module,a three-axis moving module,an ultrasonic cutter clamping module and a control module.Secondly,10 ultrasonic knives of some brand underwent fatigue testing with the table,during which non-contact measurement of the ultrasonic knife tip vibration was carried out and the Q factors were calculated at the five periods of the fatigue test,including the periods before cutting,after 500 times of cutting,after 1 000 times of cutting,after 2 000 times of cutting and after 3 000 times of cutting.Finally,the average cutting speed and burst pressure for coagulated vessels were computed at each period to validata the effectiveness of the method proposed.Results It's indicated that Q factor could effectively reflect the fatigue degradation of the ultrasonic knife tip,while the average cutting speed and burst pressure for coagulated vessels were difficult to efficiently evaluate the fatigue degradation level of the ultrasonic knife tip due to the uncertainty factors in the measurement process.Conclusion The proposed Q factor-based evaluation method can directly evaluate fatigue vibration of the ultrasonic knife tip in an accurate and quantitative manner.[Chinese Medical Equipment Journal,2024,45(6):17-22]