Maintenance and improvement program for WYD2000 field surgical lamp
10.19745/j.1003-8868.2024082
- VernacularTitle:WYD2000型野战手术灯的维修及改进方案
- Author:
Ming-Ming ZHU
1
;
Ji-Su CAO
;
Ze-Rui ZHANG
;
Jiang-Hui HAO
;
Rui-Zhe WANG
;
Feng ZHOU
;
Lu-Chuan XUN
Author Information
1. 西宁联勤保障中心药品仪器监督检验站,兰州 730050
- Keywords:
WYD2000 field surgical lamp;
troubleshooting;
improvement program
- From:
Chinese Medical Equipment Journal
2024;45(4):116-118
- CountryChina
- Language:Chinese
-
Abstract:
The steps for installation and withdrawal of WYD2000 field surgical lamp were introduced.The failures and causes of broken cross-arm connector of WYD2000 field surgical lamp were analyzed.The problems of WYD2000 field surgical lamp in vulnerability to breaking and difficulty in maintenance were solved by designing and manufacturing a special maintenance tool and optimizing the materials and fixing mode of cross-arm connection.References were provided for main-tenance and improvement of WYD2000 field surgical lamp.[Chinese Medical Equipment Journal,2024,45(4):116-118]