Optimization of batch dispensing parameters for dispensing robots
10.19745/j.1003-8868.2024026
- VernacularTitle:配药机器人批量调配参数的优化
- Author:
Hai-Wen DING
1
;
Tong TONG
;
Li-Li DU
;
Sheng LIU
Author Information
1. 中国科学技术大学附属第一医院(安徽省立医院)药学部,合肥 230001
- Keywords:
dispensing robot;
dispensing parameter;
batch dispensing;
insoluble particle
- From:
Chinese Medical Equipment Journal
2024;45(2):35-40
- CountryChina
- Language:Chinese
-
Abstract:
Objective To optimize the batch dispensing parameters of dispensing robots to ensure the standardization and homogenization of finished infusion.Methods Five parameters for intelligent dispensing robots were optimized with orthogonal test including push and pull times,standing time,needle rotation speed,solvent volume and pulling angle,with the insoluble particles as the evaluation index.Four finished infusion solutions including pantoprazole sodium solution,esomeprazole sodium solution,omeprazole sodium solution and eplazole sodium solution prepared at room temperature for 0,2,4,6 and 8 h were compared in terms of appearance,pH,osmolality and number of insoluble particles;the amounts of drug residue in empty syringe bottles and the efficiency of batch dispensing of the four finished infusion solutions using dispensing robot and manual operation were observed.Results After optimization the batch dispensing parameters of the robot were set as 2 times(push and pull times),5 s(standing time),low speed(needle rotation speed),6 mL(solvent volume)and 90°(pulling angle).After 8 h of dispensing at room temperature,the eplazole sodium solution became light yellow and clarified,and the number of insoluble particles went beyond the prescribed value in Chinese Pharmacopoeia 2020;the remained three solutions were colorless and clarified at 0,2,4,6 and 8 h,and the number of insoluble particles complied with the requirements of Chinese Pharmacopoeia 2020;there were no significant changes in pH and osmolality at 0,2,4,6 and 8 h.The robot had less drug residue than manual operation,and the efficiency of two robots operated by one person was higher than that of one person manually or one robot by one person,and all the differences were statistically significant(P<0.05).Conclusion The batch dispensing parameters optimized with orthogonal test contributes to the standardization and homogeneity of the finished solutions with decreased drug residue while increased batch dispensing efficiency.[Chinese Medical Equipment Journal,2024,45(2):35-40]