Comparison of OPD-Scan Ⅲ and IOL-Master 700 in measuring corneal curva-ture and horizontal corneal diameter before cataract surgery
10.13389/j.cnki.rao.2024.0076
- VernacularTitle:OPD-Scan Ⅲ及IOL-Master 700测量白内障术前角膜曲率和水平角膜直径的比较
- Author:
Xiaona GUO
1
;
Hui LU
;
Dou KOU
;
Lu YU
;
Yu LING
Author Information
1. 100049 北京市,航天中心医院
- Keywords:
OPD-Scan Ⅲ;
IOL-Master 700;
corneal curvature;
white-to-white distance
- From:
Recent Advances in Ophthalmology
2024;44(5):391-395
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the difference and consistency between OPD-Scan Ⅲ and IOL-Master 700 in the measurement of corneal curvature(flat keratometry,K1 and steep keratometry,K2)and horizontal corneal diameter(white-to-white(WTW)distance).Methods Totally 268 patients with 328 eyes(164 right eyes and 164 left eyes)who underwent cataract surgery at the Department of Ophthalmology of the Aerospace Centre Hospital from October 2021 to September 2022 were selected for this study.The K1,K2,and WTW values of the sampled right or left eyes were measured and analyzed using OPD-Scan Ⅲ and IOL-Master 700,respectively.Parameter comparisons were performed through paired t-tests;correlations between parameters were detected through the Pearson correlation analysis;and the Bland-Altman method and intragroup correlation coefficient(ICC)analysis were employed to determine the consistency of parameter measurements between the two instruments.Results The K1 and K2 values measured by OPD-Scan Ⅲ were greater than those measured by IOL-Master 700,while the WTW values were lower than those measured by IOL-Master 700,but the differences were statistically significant(all P<0.001).K1,K2,and WTW values measured by OPD-Scan Ⅲ were positive-ly correlated with the corresponding values measured by IOL-Master 700,and the differences were statistically significant(all P<0.001).The proportions of K1,K2,and WTW values outside the 95%limits of agreement for both instruments were within 5%,but the absolute value of the difference in values within the 95%limits of agreement was close to or more than(1.0)D,indicating a sizeable clinical deviation.ICC analysis confirmed a good consistency between K1 and K2 values of the left and right eyes measured by the two instruments(ICC>0.90).The difference in WTW values measured by the two instruments was significantly correlated with K2 values(both P<0.05).There were 5 samples(83.33%)outside the positive deviation range of WTW values,of which K2 measured by IOL-Master 700 was above 47.03.Conclusion The OPD-Scan Ⅲ and IOL-Master 700 have been found to have measurement biases when assessing K1,K2 and WTW.In clini-cal practice,the two instruments cannot be interchanged as an alternative to each other.The WTW values measured by IOL-Master 700 are greater than those obtained by OPD-Scan Ⅲ;when K2≥47.03,the WTW values may not be reliably ref-errable.