Evaluation of an Automated Instrument, PREVI Isola(R) for Inoculation of Body Fluids and Urine Samples onto Agar Plates.
- Author:
Yoonjung KIM
1
;
Seoyoung YOON
;
Young Sook SOHN
;
Yangsoon LEE
;
Hae Sun CHUNG
;
Woonhyoung LEE
;
Dongeun YONG
;
Seok Hoon JEONG
;
Kyungwon LEE
;
Yunsop CHONG
Author Information
- Publication Type:Original Article
- Keywords: Microbiological technique; Laboratory automation; Body fluids; Urine
- MeSH: Agar; Automation, Laboratory; Body Fluids; Microbiological Techniques; Sensitivity and Specificity; Stem Cells
- From:Laboratory Medicine Online 2011;1(2):105-109
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: In most clinical microbiology laboratories, inoculation of specimens on plates is performed manually and is a time-consuming process. The efficiency of this process can be improved by using an automated instrument. Currently, several automated instruments have been introduced for inoculation of samples. In this study, we have evaluated an automated instrument, PREVI Isola(R) (Biomerieux, France), used for inoculation of body fluids and urine specimens. METHODS: Both manual and automated instrument methods were used to inoculate 74 body fluid and 204 urine samples. Precision was evaluated by testing 3 types of urine samples (A, 6x10(3) colony-forming units (CFU)/mL; B, 3x10(4) CFU/mL; and C, >10(6) CFU/mL) in replicates of 20. Results of the 2 methods were compared by counting the isolated colonies on agar plates after incubation. The time required for both methods was also compared. RESULTS: The coefficient of variation (CV) of samples A, B, and C examined using the automated instrument method was 176.1%, 18.1%, and 12.6%, respectively. The sensitivity and specificity of testing body fluid samples were 77% and 100%, respectively, and those of urine samples were 87% each. The time required for testing 15 body fluid specimens and that for inoculation of each specimen was 9.7 min shorter using PREVI Isola(R) than using the manual method. CONCLUSIONS: The results of body fluid and urine culture by inoculation using the automated instrument, PREVI Isola(R), showed relative good agreement with those obtained using the manual method. The use of PREVI Isola(R) would be expected to reduce the time and labor involved in inoculating various kinds of specimens.