Application of Transimpedance Matrix Measurement in Cochlear Implantation
10.3969/j.issn.1006-7299.2024.04.018
- VernacularTitle:跨阻矩阵测量在人工耳蜗植入术中的应用研究进展
- Author:
Shujin XUE
1
;
Xingmei WEI
;
Ying KONG
;
Yongxin LI
Author Information
1. 首都医科大学附属北京同仁医院耳鼻咽喉科(北京 100730)
- Keywords:
Cochlear implant;
Transimpedance matrix;
Cochlear implant telemetry;
Electrical field im-aging
- From:
Journal of Audiology and Speech Pathology
2024;32(4):375-379
- CountryChina
- Language:Chinese
-
Abstract:
When electrical stimulation is generated by the electrode arrays of cochlear implant(CI),the cur-rent-voltage distribution in the cochlea can be recorded by the electrodes and reflected as impedance or transimped-ance.Transimpedance matrix(TIM)measurement is an objective measurement tool that utilizes the telemetry func-tion of the cochlear implant,which has shown clinical applicability in recent years for cochlear implantation.TIM of-fers convenient operation,speedy data processing,and accurate results,and has demonstrated potential application in the evaluation of electrode position intraoperatively,measurement of postoperative psychophysical loudness and assessment of neural excitation diffusion.We will introduce the principles and methods of TIM measurement and re-view the main research progress in the application of TIM measurement in cochlear implantation.