Effect of Residual Background Noise under Different Subject States on Kalman-weighted Averaging Non-sedated ABR
10.3969/j.issn.1006-7299.2024.04.005
- VernacularTitle:受试者测试状态对卡尔曼加权平均ABR测试残余噪声的影响
- Author:
Zhiyuan ZHANG
1
;
Suju WANG
;
Jiayan YANG
;
Xinheng RU
;
Daofeng NI
;
Hong JIANG
;
Yingying SHANG
Author Information
1. 北京协和医院国际医疗部(北京 100730)
- Keywords:
Auditory brainstem response;
Kalman-weighted averaging;
Residual noise;
Amplitude
- From:
Journal of Audiology and Speech Pathology
2024;32(4):313-317
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the impact of a subject's testing state on residual noise level and ABR wave V amplitude during non-sedated ABR testing using Kalman-weighted averaging(KWA).Methods Twenty-one adults(18~34 years old,42 ears)with normal hearing were enrolled for non-sedated ABR testing under three different states(lying,sitting,and writing)in a quiet room using a new Kalman-weighted averaging ABR system(vivosonic integrity system).The residual noise level and the amplitude of wave V for click ABR(cABR)of each subject were recorded.The traditional ABR test system(interacoustics,IA)was also used to record ABR with the residual noise level and the amplitude of wave V measured at the same time.Results ① There was no significant difference in am-plitude of wave V between traditional ABR and non-sedated ABR in three different testing states(P>0.05).②The residual noise levels in the lying and sitting states of KWA ABR were lower than those of traditional ABR,but there was no statistically difference(P<0.05).The residual noise level of the KWA ABR system in writing state was significantly higher than that of the other three conditions(P<0.05).③ There was no significant difference between the left and right ears in the residual noise level and amplitude of wave V for non-sedated ABR in writing state(P>0.05).Conclusion Compared with traditional ABR,the non-sedated KWA ABR system in uriting state was significantly higher than that of the other three conolitions.Haw ever,the residual noise level in lying and sit-ting states had no significant difference with conditional ABR,and different states of the subject had impact on the residual noise level in the ABR testing to sone extent.