Design and application effects of a new-type anti-slippage infusion set
10.3760/cma.j.issn.1674-2907.2017.15.026
- VernacularTitle:新型防滑脱输液器的设计与应用效果
- Author:
Zhihong ZHU
1
Author Information
1. 314000,浙江省嘉兴市第一医院感染科
- Keywords:
Diffusion of innovation;
Anti-slippage infusion set;
Meditation safety;
Infusion reaction
- From:
Chinese Journal of Modern Nursing
2017;23(15):2054-2055
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the efficacy and safety of a new-type anti-lippage infusion set. Methods Totally 400 inpatients receiving infusion between January and December 2015 were selected and equally and randomly divided into an observation group and a control group. Patients in the observation received intravenous infusion with an independently-developed and new-type anti-slippage infusion set, while patients in the control group received infusion with common disposable infusion sets. Then their accidental slippage rate and incidence of infusion reaction were observed.Results As for the control group, totally 29 patients saw accidental slippage and 9 patients showed adverse reactions to infusion,while no patients in the observation group saw accidental slippage and only 1 patients had adverse infusion reactions(χ2=31.27, 6.56;P<0.05).Conclusions The new-type anti-slippage infusion set reduces the accidental slippage rate, and it is safe and worthy of clinical application and expansion.