The influence of individualized low-dose CT imaging parameters on liver fat content measurement by quantitative CT
10.3969/j.issn.1002-1671.2024.10.032
- VernacularTitle:个体化低剂量CT成像参数对定量CT肝脏脂肪含量测量的影响
- Author:
Shiyu WANG
1
;
Yijun LIU
;
Wei WEI
;
Beibei LI
;
Xu WANG
;
Xiaoyu TONG
Author Information
1. 大连医科大学附属第一医院放射科,辽宁 大连 116011
- Keywords:
fat content;
automatic tube voltage modulation;
computed tomography;
radiation dose;
image quality
- From:
Journal of Practical Radiology
2024;40(10):1717-1721
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the influence of individualized low-dose CT imaging parameters on quantitative computed tomography(QCT)liver fat content measurement and image quality by using the combined technique of Auto-prescription and adaptive statistical iterative reconstruction(ASIR-V)algorithm with different weights.Methods A total of 231 patients who underwent both chest and abdominal CT scans were prospectively selected.The overlapping liver of two-position scans in the same case was studied,in which the chest was scanned with an individualized low-dose protocol(group A)and the upper abdomen was scanned with a routine 120 kVp protocol(group B).Patients with group A were scanned by using 80 kVp and 100 kVp based on Auto-prescription technique,and divided into the subgroups of A1 and A2;meanwhile the same patients with group B were scanned by using 120 kVp and divided into the subgroups of B1 and B2.The images were transferred to AW4.6 and QCTpro workstations for measuring CT values,standard deviation(SD)values and Fat%QCT values of the left lobe,right anterior lobe,and right posterior lobe of the liver in the overlapping regions.The signal-to-noise ratio(SNR)and contrast-to-noise ratio(CNR)were also calculated using the same-level erector spinae as background.Paired sample t-test was employed to compare the differences in Fat%QCT measurements between groups A and B.Spearman correlation analysis was conducted,and linear regression equations were established.Two observers evaluated image quality using a five-point scale and Kappa test was conducted to test inter-observer consistency.Results In group A,the Fat%QCT values showed no statistically significant difference among different weights of ASIR-V.However,the statistically significant difference was found between groups A and B(P<0.05),and there were highly positive correlations between subgroups A1 and B1 and between subgroups A2 and B2(r=0.939 7,0.987 2,P<0.05).The linear regression equation under 80 kVp and 100 kVp were as follows:y=0.976x+3.119,y=1.007x+2.041.The SNR and CNR of group A combined with post-60%ASIR-V were higher than those of group B combined with post-40%ASIR-V under conventional tube voltage.The subjective scores between groups A and B had no statistically significant difference.Conclusion Low-dose chest imaging based on Auto-prescription technique combined with post-60%ASIR-V can not only satisfy clinical diagnostic requirement,but also realize quantitative analysis of Fat%QCT under low tube voltage(80 kVp/100 kVp).