The effect of deep learning image reconstruction combined with"double-low"technique on the image quality of coronary CT angiography in overweight patients
10.3969/j.issn.1002-1671.2024.10.031
- VernacularTitle:深度学习图像重建联合"双低"技术对超重患者冠状动脉CT血管成像图像质量的影响
- Author:
Li SHEN
1
;
Hui PENG
;
Zhanli REN
;
Nan YU
;
Dong HAN
;
Tao QIN
;
Yongjun JIA
;
Yuxin LEI
;
Yangyang YAN
Author Information
1. 陕西中医药大学附属医院影像科,陕西 咸阳 712000
- Keywords:
coronary computed tomography angiography;
deep learning image reconstruction;
iterative reconstruction;
image quality
- From:
Journal of Practical Radiology
2024;40(10):1712-1716
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the effect of deep learning image reconstruction(DLIR)algorithm combined with"double low"[low voltage(kV)and low contrast agent dosage]technique on the image quality of coronary computed tomography angiography(CCTA)in overweight patients compared with adaptive statistical iterative reconstruction(ASIR-V)and filtered back projection(FBP).Methods Fifty-two patients with body mass index(BMI)between 25.1 kg/m2and 28 kg/m2 who underwent CCTA scanning were prospectively selected,all of whom scanned on a GE Revolution APEX-CT with a tube voltage of 80 kV,a smart mA(500-1 300 mA),a noise index of 30,and a contrast dosage of 0.5 mL/kg;four groups of images were reconstructed for each patient,FBP,50%ASIR-V,DLIR-M,and DLIR-H.The CT and SD values of the aorta(AO)root,the proximal segment of the right coronary artery(RCA),the left circumflex(LCX),the left anterior descending branch(LAD)and the pericardial fat were measured,and the signal-to-noise ratio(SNR)and contrast-to-noise ratio(CNR)were calculated.Two doctors experienced in the diagnosis of cardiovascular disease were selected to subjectively score the reconstructed images using a double-blind method.The SD value,SNR value,CNR value and subjective scores of images in the four groups were compared.Results In the objective image quality evaluation,there were statistically significant differences in SD value,SNR value and CNR value of reconstructed images in the four groups(P<0.05).As the four groups of reconstruction algorithms FBP,50%ASIR-V,DLIR-M,and DLIR-H were changed sequentially,the image SD value gradually decreased,the SNR value and CNR value gradually increased,and the DLIR-H group had the lowest SD value and the highest SNR and CNR values.In the subjective image quality evaluation,the subjective scores of the two doctors had good consistency(Kappa value=0.900),and the difference between them was statistically significant(P<0.001).The subjective scores of DLIR-M and DLIR-H groups were higher.Conclusion DLIR algorithm combined with"double-low"technique can significantly improve the CCTA image quality of overweight patients,which is better than 50%ASIR-V and FBP.