Feasibility study of reducing radiation dose of enhanced abdominal examination in 320-row CT using 80 kV tube voltage combined with Karl iterative reconstruction
10.3969/j.issn.1002-1671.2024.07.030
- VernacularTitle:联合80kV管电压及Karl迭代重建降低320排CT腹部增强扫描辐射剂量的可行性研究
- Author:
Jianxin CAO
1
;
Zhuo ZHU
;
Wei XU
;
Jiashuai LI
;
Qing ZHANG
Author Information
1. 武汉市普仁医院放射科,湖北 武汉 430081
- Keywords:
tube voltage;
iterative reconstruction;
computed tomography;
radiation dose
- From:
Journal of Practical Radiology
2024;40(7):1170-1174
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the feasibility of decreasing radiation dose of enhanced abdominal examination in 320-row CT using 80 kV tube voltage combined with Karl iterative reconstruction.Methods One hundred patients with body mass index(BMI)<25.0 kg/m2 who underwent enhanced abdominal examination in 320-row CT were divided into group A and group B(50 patients for each group).120 kV tube voltage and filtered back projection(FBP)reconstruction were used for group A,while 80 kV tube voltage,FBP recon-struction and Karl iterative reconstruction were used for group B.Radiation dose and image quality between the two groups were evaluated.Results The effective dose(ED)were(11.0±1.7)mSv in group A and(5.6±0.7)mSv in group B for single scan(P<0.01).Compared with group A,the ED in group B was decreased by 49.1%.The noise on the Karl images in group B was significantly decreased compared with the FBP images in group B(P<0.01),and there was no significant difference in noise of liver between the FBP ima-ges in group A and the Karl images in group B(P>0.05).Compared with the FBP images in group A,the noise of abdominal aorta and portal vein on the Karl images in group B was slightly increased(P<0.05),while the signal-to-noise ratio(SNR)and the con-trast-to-noise ratio(CNR)of abdominal aorta and portal vein on the Karl images in group B were significantly increased(P<0.01).Image quality scores for liver,spleen,pancreas,kidney,gastrointestinal tract,abdominal aorta and portal vein on the Karl images in group B were significantly higher than those on the FBP images in group B(P<0.01),and there was no significant difference in the image quality scores for liver,spleen,pancreas,kidney,gastrointestinal tract,and abdominal aorta between the FBP images in group A and the Karl images in group B(P>0.05),while the image quality score for portal vein on the Karl images in group B was higher than that on the FBP images in group A(P<0.05).Conclusion Radiation dose can be decreased obviously in 320-row CT enhanced abdominal examination using 80 kV tube voltage combined with Karl iterative reconstruction for the patients with BMI<25.0 kg/m2,while image quality can be maintained for diagnosis.