Performance Evaluation of a Point-of-care Test, ‘Samsung LABGEO PA CHF Test’, for the Amino-terminal Pro-brain Natriuretic Peptide.
- Author:
Man Jin KIM
1
;
Kyunghoon LEE
;
Sun Hee JUN
;
Sang Hoon SONG
;
Woon Heung SONG
;
Junghan SONG
Author Information
- Publication Type:Original Article
- Keywords: NT-proBNP; Point-of-care testing; Performance evaluation
- MeSH: Adult; Diagnosis; Emergencies; Healthy Volunteers; Heart Failure; Humans; Methods; Plasma; Point-of-Care Systems*; Point-of-Care Testing
- From:Laboratory Medicine Online 2017;7(3):135-140
- CountryRepublic of Korea
- Language:Korean
- Abstract: BACKGROUND: The amino-terminal pro-brain natriuretic peptide (NT-proBNP) is a useful biomarker for the diagnosis of acute congestive heart failure. A point-of-care test (POCT) could rapidly detect the presence of NT-proBNP during emergencies. We evaluated the analytical performance of the new Samsung LABGEO PA CHF Test (Samsung Electronics, Korea). METHODS: Based on the guidelines of the Clinical and Laboratory Standards Institute (CLSI), we compared the precision, linearity, and method with those of the E170 (Roche Diagnostics, Switzerland). Matrix comparison between the NT-proBNP values in whole blood and plasma was also performed, and the reference interval was determined using residual samples from healthy adults selected based on the evaluation criteria. RESULTS: The Samsung LABGEO PA CHF Test provided results in approximately 18 min. The coefficient of variation (CV) of within-laboratory precision was below 6.8%. A desirable linearity was observed in the range of 0–10,000 pg/mL, with R²=0.99. The correlation with E170 was also excellent (N=108, r=0.96). NT-proBNP values in the whole blood were correlated with those in the plasma (N=36, r=0.99). The reference interval for the circulating NT-proBNP concentration was determined in 118 plasma samples from healthy subjects (26-75 yr of age). The 97.5th percentile was found to be 58.3 pg/mL. CONCLUSIONS: The Samsung LABGEO PA CHF Test demonstrated a good analytical performance. It could be a powerful tool as a POCT for clinical practice, particularly during emergencies.