Application and nursing of Icare rebound tonometer
10.3760/cma.j.issn.1674-2907.2011.16.016
- VernacularTitle:Icare回弹式眼压计的应用与护理
- Author:
Min WANG
1
;
Haiyan QIAN
;
Fang-Li LIN
;
Ye GUAN
;
Xiuyun CHEN
Author Information
1. 解放军总医院
- Keywords:
Tonometry,ocular;
Nursing;
Icare rebound tonometer
- From:
Chinese Journal of Modern Nursing
2011;17(16):1901-1902
- CountryChina
- Language:Chinese
-
Abstract:
Objective To observe the clinical application and nursing of Icare rebound tonometer.Methods Intraocular pressure (IOP) measurements were performed on 112 cases (112 eyes) with ICARE rebound tonometer, measure on central cornea 2mm distance to temporal and nasal limbus corneae respectively.Results The results of intraocular pressure measurement on central cornea, 2mm distance to temporal and nasal limbus corneae are (15.4±3.4),(15.8±3.6),(15.6±3.4)mm Hg. There were no significant difference among them(F=0.325,P>0.05), but they have a good consistency.Conclusions There was good consistency on three different position on IOP measurements. The nursing is important for correct measurement.