Tooth Design of Ophthalmic Microsurgical Forceps Based on Minimum Slip Force
10.12455/j.issn.1671-7104.230715
- VernacularTitle:基于最小滑脱力的眼科显微手术镊的齿型设计
- Author:
Tianfeng LENG
1
;
Yu ZHENG
;
Shaofeng HAN
;
Chenhan GUANG
;
Yang YANG
Author Information
1. 北京航空航天大学机械工程及自动化学院,北京市,100083
- Keywords:
microsurgical forceps;
simulation;
clamping force;
slip force;
stress
- From:
Chinese Journal of Medical Instrumentation
2024;48(6):678-682
- CountryChina
- Language:Chinese
-
Abstract:
This study proposed the use of the minimum slip force as an indicator to measure the clamping ability of microsurgical forceps.By analyzing the minimum slip forces of four typical tooth types of microsurgical forceps,the clamping capacity of each type was evaluated.Among the existing tooth forms,the staggered tooth type exhibited a relatively large minimum slip force.Consequently,a new tooth-shaped structure for microsurgical forceps-the hemispherical convex structure was proposed.Simulation and experimental studies demonstrated that this new dental structure could achieve stable and reliable clamping.