Study of Imitation Hardened Plaque RF Ablation with Different Frequencies
10.3969/j.issn.1671-7104.2017.04.002
- VernacularTitle:不同频率射频用于仿硬化斑块消融方法研究
- Author:
Yuntao MA
1
;
Jincheng ZOU
;
Aili ZHANG
Author Information
1. 上海交通大学生物医学工程学院Med-X研究院
- Keywords:
atherosclerosis;
radiofrequency;
ablation
- From:
Chinese Journal of Medical Instrumentation
2017;41(4):240-243
- CountryChina
- Language:Chinese
-
Abstract:
In order to solve the heat transfer mechanism in radiofrequency treatment of atherosclerosis, we design two experimental structures to simulate the plaque, and use three frequencies and a variety of electrode arrangement, to study the heat transfer mechanism and heating capacity under the special structure. The results show that the temperature increment and heating area increase with increasing frequency. Under the structure in which the lipid particles are embedded in the fiber cap, using three electrodes with opposite polar arrangement between middle electrode and others, achieves effective heating to lipid pool in the plaque.