Application of Image Plate in Quality Assurance of Medical Accelerator
10.3969/j.issn.1671-7104.2016.06.018
- VernacularTitle:IP板在医用直线加速器质量控制中的应用
- Author:
Guangbo LIU
1
;
Huijuan YAN
;
Yunchuan SUN
Author Information
1. 河北省沧州中西医结合医院放疗科
- Keywords:
computed radiograpy;
autoradiography film;
quality control
- From:
Chinese Journal of Medical Instrumentation
2016;40(6):458-460
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the value of Computed Radiography(CR) in quality control for the medical linear accelerator. Methods By using both CR and autoradiography film respectively, we tested the alignment of light fields with radiation fields, colimator rotation with center, couch rotation with center and multi-leave colimators(MLC) position accuracy of medical linear accelerator. Each tests were carried out ten times repetitive colection. Then compared the differences between this two methods.Results There were no significant difference between CR and autoradiography film in the same grup(P>0.05).Conclusion CR can be used in quality control for the medical linear accelerator since it can meet quality control requirements of radiotherapy.