Bone Tomography Scan Optimization in SPECT/CT Equipment of Nuclear Medicine
10.3969/j.issn.1671-7104.2016.05.017
- VernacularTitle:核医学SPECT/CT骨断层扫描方案的优化
- Author:
Yuankai WANG
1
;
Huiqing ZHU
;
Miao LIU
;
Congjin LIU
;
Xingdang LIU
Author Information
1. 复旦大学附属华山医院核医学科
- From:
Chinese Journal of Medical Instrumentation
2016;40(5):377-379
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study and search for a balance between the image quality and acquisition speed in tomography of whole body bone scan.Methods Adjustments of acquisition conditions were carried out gradualy every two months since April 2014. The qualities of fused SPECT/CT images were diagnosed by three doctors. Then the picture would be evaluated comprehensively by analyzing image quality and image resolution after adjusting image acquisition conditions.Results Seven kinds of image acquisition conditions taken were in line with diagnostic requirements. The third method is extended to clinical work best.Conclusion To obtain a high colection effi ciency, parameters of bone tomography acquisition can be set a frame of 5 seconds, total 64 (5.625o), automatic probe close and continuous scanning. Also recommends the use of "continuous" instead of "step and shoot" approach in bone SPECT acquisition. tomography, emission-computed, single-photon, bone tomography, program optimization.