Evaluation of the EDOSE for IMRT Delivery Quality Assurance
10.3969/j.issn.1671-7104.2016.02.022
- VernacularTitle:EDOSE进行调强剂量验证结果的评估
- Author:
Xiaoshen WANG
1
;
Zhongjian JU
;
Wei XU
;
Hanshun GONG
;
Zhongkai JIANG
;
Xiaohu CONG
Author Information
1. 中国人民解放军总医院 放疗科
- Keywords:
EDOSE;
EPID;
delivery quality assurance
- From:
Chinese Journal of Medical Instrumentation
2016;40(2):147-149
- CountryChina
- Language:Chinese
-
Abstract:
Objective Evaluation of intensity-modulated radiation therapy(IMRT) delivery quality assurance(DQA) using EDOSE software based on accelerator electronic portal imaging device (EPID).Methods Analysed the dose response uniformity correction of EPID imaging plate in different fields and the gamma pass rates about the fields; And compared the calculated parameters of the EDOSE and the Pinnacle planning station.Results The uniformities on the homogeneous region of thefi elds are good after who corrected by EDOSE. The gamma(γ, 3mm/3%) pass rates are in the ideal range. The volume of the organs at risk(OAR) have no statistical signifi cance except the right lung; A part of the parameters have statistical signifi cance.Conclusions The results of the calculation and analysis of EDOSE has a high agreement with the planning system.