Infant Incubator Temperature Monitoring Difference Analysis and Research
10.3969/j.issn.1671-7104.2016.02.016
- VernacularTitle:婴儿培养箱温度监测差异分析和研究
- Author:
Xi WANG
1
;
Zhi ZHUO
;
Hongjia FU
;
Gang DENG
;
Bo GAO
Author Information
1. 四川大学华西第二医院
- Keywords:
infant incubator;
temperature;
temperature monitoring;
the temperature deviation
- From:
Chinese Journal of Medical Instrumentation
2016;40(2):128-130
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss the temperature difference of infant incubator treatment for the baby to provide important guarantee, to ensure its safe operationMethods Using a completely independent of the infant incubator's temperature monitoring and alarm system of infant incubator temperature and monitoring and alarm system, 20 sets of real-time monitoring and automatic logging data, different brand infant incubator temperature changes compared with infant incubator's own body temperature, and analysis.Results20 sets of different brand infant incubator, 10(50%) in the devices panel display data with a monitoring device differences in measured data, clinical pose a safety hazard. Conclusions For clinical use of infant incubator temperature real-time monitor, by the monitoring system for additional auxiliary monitoring and alarm, thereby improve infant incubator clinical application security, reduce the incidence of related medical accidents, improve the quality of medical treatment.