Measurement Error Analysis and Calibration Technique of NTC- Based Body Temperature Sensor
10.3969/j.issn.1671-7104.2015.06.002
- VernacularTitle:一种基于NTC的体温传感器测量误差分析及校准技术
- Author:
Chi DENG
1
;
Wei HU
;
Shengxi DIAO
;
Fujiang LIN
;
Dahong QIAN
Author Information
1. 中国科学技术大学电子科学与技术系
- Keywords:
temperature measurement;
NTC thermistor;
wearable;
temperature error calibration[Key words
- From:
Chinese Journal of Medical Instrumentation
2015;(6):395-399
- CountryChina
- Language:Chinese
-
Abstract:
A NTC thermistor-based wearable body temperature sensor was designed. This paper described the design principles and realization method of the NTC-based body temperature sensor. In this paper the temperature measurement error sources of the body temperature sensor were analyzed in detail. The automatic measurement and calibration method of ADC error was given. The results showed that the measurement accuracy of calibrated body temperature sensor is better than±0.04oC. The temperature sensor has high accuracy, smal size and low power consumption advantages.