The Study of the Measurement of Heart Rate Variability Using ECG and Photoplethysmographic Signal
10.3969/j.issn.1671-7104.2015.04.004
- VernacularTitle:心电图和光电容积脉搏波计算心率变异性的比较研究
- Author:
Buqing WANG
1
;
Xiaoke CHAI
;
Zhengbo ZHANG
;
Weidong WANG
Author Information
1. 解放军总医院医学工程保障中心
- Keywords:
spontaneous breathing;
ECG;
photoplethysmographic signal;
heart rate variability
- From:
Chinese Journal of Medical Instrumentation
2015;(4):249-252,264
- CountryChina
- Language:Chinese
-
Abstract:
In comparison with the measurement of heart rate variability from ECG and from photoplethysmographic signal from 46 healthy adults in their spontaneous breathing state. The beat-to-beat intervals in ECG and pulse-to-pulse intervals in photoplethysmographic signal are extracted,and then the parameters of heart rate variability are calculated. Three kinds of algorithms are chosen to get the pulse-to-pulse intervals, which are the intervals of maximum of second derivative,the maximum of PPG signal and the tangent intersection. The results show that the correlation coefficients of the HRV parameters in the two calculation methods are highly correlated. The Bland-Altman scattered plots show the relative bias results from the algorithm of the maximum of PPG signal are smal est and singular points that deviate from the consistent limits are the least compared with the other two algorithms.