Abstract:
Objective A method for determined oxygen permeability(Dk) of finished soft contact lens. Methods Oxygen permeability(Dk) of 10 group finished soft contact lens was determined by polarographic method, and the values were linearly calibrated using the measured and established Dk values of the specified reference materials. Results Only one Dk value fel outside of the requirement of GB 11417.3—2012 tolerance for Dk(±20%). Conclusion It is great practicability when determined the oxygen permeability(Dk) of finished soft contact lens by polarographic method which were linearly calibrated using the measured and established Dk values of the specified equal thickness reference materials. without the measured lens customized 4 different thickness.