The Design of Bionic Left Ventricular Auxiliary Pump
10.3969/j.issn.1671-7104.2015.01.008
- VernacularTitle:仿生左心室辅助血泵设计
- Author:
Henglin JIN
1
;
Xiaobing HU
;
Lei DU
Author Information
1. 四川大学制造科学与工程学院
- Keywords:
artificial heart;
left ventricular auxiliary pump;
bionic design;
animal experiment
- From:
Chinese Journal of Medical Instrumentation
2015;(1):29-32
- CountryChina
- Language:Chinese
-
Abstract:
This paper reports a novel design of bionic left ventricular auxiliary pump, and the characteristic is that elastic diaphragm of pump driven by hydraulic, having smooth, reliable blood supply, can prevent blood clots, can use the flow sensor, pressure sensor detection showing the blood pressure and blood volume at the inlet and outlet of the pump. The pump can go with heart rate synchronization or asynchronous auxiliary by the R wave of human body’s ECG. The design goal is realization of bionic throb. Through the animal experiment, the blood pressure waveforms are close to expectations, stable flow can stroke according to the set value, which prove that the pump can meet the requirement for heart disease patients for bionic left ventricular assistant.