Fluid Solid Interaction Analysis of Bioprosthetic Heart Valve
10.3969/j.issn.1671-7104.2014.05.004
- VernacularTitle:人工生物瓣膜流固耦合分析
- Author:
Xuejie MA
1
;
Yawei DU
;
Linan ZHANG
;
Zengtao HOU
;
Xin YE
Author Information
1. 广东科学技术职业学院
- Keywords:
bioprosthetic heart valve;
mechanical property analysis;
fluid solid interaction analysis
- From:
Chinese Journal of Medical Instrumentation
2014;(5):325-328
- CountryChina
- Language:Chinese
-
Abstract:
This paper constructs numerical models of bioprosthetic heart valve and blood. Thefluid solid interaction is carried out using penalty function method. The mechanical property of the bioprosthetic heart valve during cardiac cycle is simulated with ANSYS software. Results show that the Von Mises stress concentrates at the junction of attachment edge and coaptation edge. The open time of bioprosthetic heart valve is consistent with that of actural measurement. The peak velocity of blood is in the range of physiology. This model provides more realistic mechanical property of bioprosthetic heart valve during cardiac cycle compared to pure solid model, and facilitates design and optimization of bioprosthetic heart valve.