Discussion on Testing of Flow Rate of Infusion Device about Industry Standard
10.3969/j.issn.1671-7104.2014.03.017
- VernacularTitle:关于行业标准中输注泵流量检测的探讨
- Author:
Songhe HUA
1
Author Information
1. 上海市医疗器械检测所
- Keywords:
infusion device;
industry standard;
fl ow rate;
dangerous coeffi cient
- From:
Chinese Journal of Medical Instrumentation
2014;(4):292-294
- CountryChina
- Language:Chinese
-
Abstract:
Carried on the exploration testing offl ow rate of infusion device about industry standard YY 0451-2010. Engaged in fl ow rate experiments adopting different method that are provided by new and old industry standard for samples of the same type. Compared with the result of the dangerous coeffi cient by calculating the test data, the old standard can be more sensitive to refl ect the situation of productfl ow rate, so it can be applied to conventional control of the products. The method which provided by the new industry standard is suitable for evaluating periodicity the level of product contaminated.